Type |
Semi-automatic,
single-beam filter photometer
|
| Light
Source |
Halogen
lamp - 12 V, 20 W with
protection for lifetime
|
| Wavelength |
340
nm - 800 nm |
| Wavelength
Selection |
Automatic
via 9-position filter
wheel :
6 standard interference
filters: 340 nm, 405 nm,
492 nm, 546 nm, 578 nm
and 623 nm;
3 positions for optional
filter of choice |
| Photometric
Range |
0
- 2.5 A |
| Cuvette
System |
Micro
flow cell: 32 µl, 10 mm
light path interchangeable
with normal standard cuvettes
(macro or semi-micro,
disposable or special
optical glass) |
| Temperature
Control |
- Internal
Peltier element, temperature
variable, pre-adjusted
to 25 °C, 30 °C and
37 °C
- Equilibration
time for aspirated
reaction mixture to
reach 37 °C from ambient
temperature: 15 sec
|
| Aspiration
System |
Built-in
peristaltic pump driven
by stepper motor, programmable
aspiration volume controlled
by infrared light barrier
|
| Sipping
Volume |
- Minimum
250 µl, typically
500 µl up to 2000
µl
- Separate
setting of aspirate
volume and wash volume
|
| Operator
Interface |
Touchscreen
for direct functions and
alphanumerical inputs
|
| Data
Presentation |
Graphic
display: white characters
or symbols, blue backlight,
dimension 240 * 128 dots
|
| Integrated
Printer |
Dot-matrix
printer, 24 characters
per line |
| Languages
|
English
and German / Indonesian
/ Russian / Spanish |
| Memory
|
- General
operating software
can be updated by
PC
- Reagent
open system with capacity
for up to 231 programmable
methods
- Import
of data by touchscreen
or PC
- Up
to 50 non-linear calibration
curves with max 20
sets of points can
be stored
|
| Data
Logging |
Up
to 2970 results of max
99 samples can be managed
in the memory |
| Signal
Port |
Serial
port for connection to
an external printer and/or
PC |
| Measurement
Procedures |
- Absorbance
- End
Point with factor,
standard or multiple
standards, with or
without reagent blank
and/or sample blank
- Bichromatic
end point
- Kinetics
with standard, factor
or multiple standards,
with or without blank
- Fixed
time with standard,
factor or multiple
standards, with or
without reagent blank
- Turbidimetry
- Single,
double and triple
determinations
- Curve
fitting for non-linear
standard curves
- Free
hemoglobin in combination
with optional interference
filters
|
| Quality
control |
Up
to 50 methods can be controlled
with two control sera
Levey Jennings plot |
| Measuring
Time |
- Kinetic:
variable from 3 -
25 deltas, time per
delta 4 - 255 sec
in 1 sec steps
- Fixed
time: variable from
0 - 65535 sec in 1
sec steps
|
| Delay
Time |
Programmable
from 0 - 65535 sec in
1 sec steps |
| Mains
Supply |
Range:
90 VAC up to 264 VAC at
50/60 Hz |
| Dimensions
|
Length
33 cm x width 34 cm x
height 18 cm |
| Weight
|
6
kg |